Advanced DSC Processing

How is the arterial input function used to extract more quantitative flow information from the DSC data?
DSC deconvolution
Conversion of raw DSC imaging data into quantitative blood flow measurements takes place in three steps: 
  1. Translation of tissue MR signal s(t) into gadolinium concentration c(t); 
  2. Selection of an appropriate arterial input function a(t); and 
  3. Mathematical modeling and computation. 
The first two steps have been described in prior Q&As. This section addresses mathematical methods to quantify blood flow knowing only c(t) and a(t). The analysis comes from from linear systems theory and requires several important assumptions:
  1. Blood flow (F) and blood volume (V) remain constant during the measurement period.
  2. All injected tracer molecules eventually flow out of the system. In other words, there is no permanent trapping, pooling, or loss of tracer within the tissue.
  3. The system has a linear response to inputs and follows the principle of superposition. This means that the total output response from two independent inputs is the same as the sum of responses from those inputs considered separately.
The reader will immediately recognize how easily this simple model may fail in practice. For example, a brain tumor accumulating gadolinium contrast through a leaky blood-brain-barrier would not meet assumption #2. 
Blood volume (V) is the first and easiest parameter to extract. Assuming no leakage or recirculation effects, V is proportional to the area under the tissue concentration curve c(t) during the first pass of contrast. Theoretically this relationship can be made quantitative by dividing by the gadolinium concentration a voxel containing only blood. Typically such a reference voxel is chosen to lie within a feeding artery, and the arterial input function curve, a(t), is convenient to use. (A voxel from a large vein can also be chosen, a common practice in CT perfusion studies).  In mathematical form 
blood volume equation dsc
Blood flow (F) is the next parameter to be calculated, but the process for doing so is somewhat "convoluted". (Please forgive the old physics pun). Understanding the method requires a brief introduction to linear systems theory.
Dirac and residue function
A linear system (e.g., the tissue capillary bed) can be analyzed by considering its response to a hypothetical infinitely high intensity, infinitely short duration impulse called the Dirac delta function δ(t). For our system δ(t) could be viewed as the instantaneous injection of a bolus of contrast particles into the tissue through an arterial port at time t=0.  After injection, some of these particles will find relatively direct pathways to the venous port; others will choose more circuitous routes and take longer to traverse the capillary bed. This results in a dispersion of the injected contrast bolus within tissue and a range of particle transit times. 

The fraction of injected contrast particles remaining in the tissue at time t after such an impulse injection is known as the residue, described by the dimensionless function R(t). Immediately after injection (before any particles have left the system) R(t) is at its maximum with value R(0) = 1. As particles flow out of the system, R(t) decreases, eventually reaching final value R(∞) = 0 when no contrast remains in the tissue.
The width of the residue function reflects the distribution of particle transit times through a tissue. Tissues with rapid transit times have sharp, narrow residue curves, while tissues with prolonged transit times have wide ones. The area under the residue function reflects the average time a particle spends transitioning through the tissue vascular bed, an important physiological parameter known as the mean transit time (MTT). In mathematical form
Picture
Picture
Upper figures: a real-life arterial input function (AIF) with corresponding resultant tissue concentration curve c(t)
Lower figures: modeling the tissue response by a set of Dirac delta functions
In real life it is impossible to deliver an arterial bolus of contrast to tissue instantaneously. The arterial input function, a(t), is thus always dispersed in time as shown in the figure. This broad arterial input, however, can be represented as a set of Dirac delta functions at different time delays (τ), each producing an independent response. The mathematical operation of scaling and summing these individual impulse responses to generate the tissue contrast concentration curve c(t) is known as convolution. In mathematical notation
convolution equation
where F is blood flow and the "circle/x" symbol is the convolution operator. The convolution process sums together ("integrates") the individual arterial inputs weighted ("multiplied") by the residue function at various time delays (τ) after injection. To account for the fact that gadolinium contrast distributes only in the extracellular space of blood vessels, some versions of this equation contain a constant factor (ρ/h) preceding F, where ρ is the tissue density in g/mL and h is a fraction accounting for the difference in hematocrit between capillaries and larger vessels.
In a DSC perfusion experiment we measure a(t) and c(t) directly, but do not know either the blood flow (F) nor the residue function, R(t). Because there are more unknowns than knowns in this equation, a unique solution does not exist. Nevertheless, various mathematical techniques can extract reasonable estimates for these parameters, especially blood flow (F).
Three general strategies for performing mathematical deconvolution exist. The most common method converts the equation into matrix form and perform iterative calculations with a procedure known as singular value decomposition (SVD). A second approach, less widely used, is called "parametric", because it assumes a predetermined shape for the residue function such as a decaying exponential. Other deconvolution techniques involve Fourier transformation or stochastic estimation. 
The end result of all the deconvolution methods is a quantitative estimate for blood flow (F). Since blood volume (V) has already been calculated, the final step is to compute the mean transit time (MTT) on a voxel-by-voxel basis using the central volume theorem (MTT = V/F). 
It should be noted that significantly different results may be obtained depending on the choice of the AIF as well as which commercial software product for calculation is used. Even software claiming to use identical methods may give different results when applied to the same raw data. Caution is advised in relying too heavily on the absolute numbers obtained from such quantitative methods. 
The concept of (de)convolution is admittedly confusing for those without a strong mathematical background, but my colleague Josh Shimony makes it quite understandable in this vimeo video beginning at about minute 33:00. The whole video is a great introduction to both DSC and ASL, so if you have an extra hour I suggest you watch it in its entirety.

Josh Shimony: MRI Perfusion from NRG on Vimeo.

Advanced Discussion (show/hide)»

Fourier methods for deconvolution begin by taking the Fourier Transform (FT) of each side of the convolution equation for blood flow. This relies on the well-known Convolution Theorem that the Fourier Transform of a product of two functions is the convolution between the Fourier Transforms of each function. In mathematical terms

F • R(t) = FT−1{FT[c(t)] / FT[a(t)]}

Noting that the initial value of the residue function is 1 (i.e. R(0) = 1), the magnitude of the first time point of the solution equals blood flow (F). Although conceptually straightforward, this method is not frequently used because of instability and noise issues.

Many variations of the most popular technique, singular value decomposition (SVD) exist. SVD is a factorization method commonly used in matrix algebra to solve inverse problems. Because SVD is potentially ill-posed (unstable) due to round-off errors and noise, some constraints (regularization) must be applied. The most straightforward regularization method (truncation) treats all singular matrix values below a certain level as zeros. A second method (Tikhonov regularization) stabilizes the algorithm by manipulating the diagonal elements, resulting in a smoother truncation of singular values.

Another problem with the standard SVD method is that its results are sensitive to delays in contrast arrival between different parts of the tissue. Variations such as reformulated SVD and block-recirculant SVD largely overcome this limitation and are gradually being implemented by vendors in their software.

Several newer statistical methods focus more on improving estimations of the residue function and reducing its oscillations. Which method, if any, will prevail is not easily predictable at present
.

Tmax and Why I Don't Like It

A number of clinical trials in the stroke literature have used a calculated perfusion metric known as Tmax. Tmax is defined to be the time at which the residue function, R(t), calculated from deconvolution, reaches its maximum value. In theory, therefore, Tmax is the time between the peak of the arterial input function (AIF) and the peak of the residue function.

Tmax is not the same as MTT. MTT reflects the transit of contrast through the capillary bed once it arrives at the tissue of interest. Conversely, Tmax reflects the time delay and dispersion of contrast between where the AIF is measured en route to the capillary system.

Tmax is thus highly dependent on the chosen location of the AIF. The farther away from the tissue of interest, the larger will be the Tmax value calculated. Additionally, regularization (filtering) of the SVD matrix to reduce noise effects in deconvolution will alter R(t) by inducing oscillations and peak shifting, also affecting Tmax.

Notwithstanding these methodological issues, many stroke neurologists seem to love Tmax and have recommended its use in a number of clinical trials. One reason I believe it is so favored is that Tmax maps display relatively "flat" image contrast in normal regions of brain but dramatic differences in regions with poor or marginal perfusion.

In summary, Tmax does provide some information that is complementary to MTT and CBF. However, it is a "messy" parameter that reflects primarily contrast delay and dispersion between the site chosen for AIF measurement and the tissue of interest. As such, it is very AIF-dependent and should be used with caution in comparing different patients in stroke research protocols.


References
     Boxerman JL, Rosen BR, Weisskoff RM. Signal-to-noise analysis of cerebral blood volume maps from dynamic NMR imaging studies. J Magn Reson Imaging 1997; 7:528-37.
     Calamante F, Christensen S, Desmond PM, et al. The physiological significance of the time-to-maximum (Tmax) parameter in perfusion MRI. Stroke 2010; 41:1169-1174. (See Advanced Discussion tab for why I don't favor this parameter in practice.)
     Fieselmann A, Kowarschik M, Ganguly A, et al. Deconvolution-based CT and MR brain perfusion measurement: theoretical model revisited and practical implementation details. Int J Biomed Imag 2011; doi:10/1155/2011/467563
     Orsingher L, Piccinini S, Crisi G. Differences in dynamic susceptibility contrast MR perfusion maps generated by different methods implemented in commercial software. J Comput Assist Tomogr 2014; 38:647-654. (different software gives different results, even when methods are said to be the same!)
     Østergaard L, Weisskoff RM, Chesler DA, et al. High resolution measurement of cerebral blood flow using intravascular tracer bolus passages. Part I. Mathematical approach and statistical analysis. Magn Reson Med 1996; 36:715-725. (good description of deconvolution techniques including Fourier and SVD).
     Zaharchuk G. Theoretical basis of hemodynamic MR imaging techniques to measure cerebral blood volume, cerebral blood flow, and permeability. AJNR Am J Neuroradiol 2007; 28:1850-8.

Related Questions
     What is the arterial input function (AIF) and why do you need to measure it? 
     How are parameters like BF, BV, and MTT defined and used in perfusion imaging?  

←  Previous Question
Next Question  →
↑ Complete List of Questions ↑
© 2024 AD Elster, ELSTER LLC
All rights reserved.   
MRIquestions.com - Home
Donate
Please help keep this site free for everyone in the world!
  • Basic Electromagnetism >
    • What causes magnetism?
    • What is a Tesla?
    • Who was Tesla?
    • What is a Gauss?
    • How strong is 3.0T?
    • What is a gradient?
    • Aren't gradients coils?
    • What is susceptibility?
    • How to levitate a frog?
    • What is ferromagnetism?
    • Superparamagnetism?
  • Magnets - Part I >
    • Types of magnets?
    • Brands of scanners?
    • Which way does field point?
    • Which is the north pole?
    • Low v mid v high field?
    • Advantages to low-field?
    • Disadvantages?
    • What is homogeneity?
    • Why homogeneity?
    • Why shimming?
    • Passive shimming?
    • Active shimming?
  • Magnets - Part II >
    • Superconductivity?
    • Perpetual motion?
    • How to ramp?
    • Superconductive design?
    • Room Temp supercon?
    • Liquid helium use?
    • What is a quench?
    • Is field ever turned off?
    • Emergency stop button?
  • Gradients >
    • Gradient coils?
    • How do z-gradients work?
    • X- and Y- gradients?
    • Open scanner gradients?
    • Eddy current problems?
    • Active shielded gradients?
    • Active shield confusion?
    • What is pre-emphasis?
    • Gradient heating?
    • Gradient specifications?
    • Gradient linearity?
  • RF & Coils >
    • Many kinds of coils?
    • Radiofrequency waves?
    • Phase v frequency?
    • RF Coil function(s)?
    • RF-transmit coils?
    • LP vs CP (Quadrature)?
    • Multi-transmit RF?
    • Receive-only coils?
    • Array coils?
    • AIR Coils?
  • Site Planning >
    • MR system layout?
    • What are fringe fields?
    • How to reduce fringe?
    • Magnetic shielding?
    • Need for vibration testing?
    • What's that noise?
    • Why RF Shielding?
    • Wires/tubes thru wall?
  • Overview >
    • ACR Safety Zones?
    • MR safety screening?
    • Incomplete screening?
    • Passive v active implants?
    • Conditional implants?
    • Common safety issues?
    • Projectiles?
    • Metal detectors?
    • Pregnant patients?
    • Postop, ER & ICU patients?
    • Temperature monitoring?
    • Orbital foreign bodies?
    • Bullets and shrapnel?
  • Static Fields >
    • "Dangerous" metals?
    • "Safe" metals?
    • Magnetizing metal?
    • Object shape?
    • Forces on metal?
    • Most dangerous place?
    • Force/torque testing?
    • Static field bioeffects?
    • Dizziness/Vertigo?
    • Flickering lights?
    • Metallic taste?
  • RF Fields >
    • RF safety overview?
    • RF biological effects?
    • What is SAR?
    • SAR limits?
    • Operating modes?
    • How to reduce SAR?
    • RF burns?
    • Estimate implant heating?
    • SED vs SAR?
    • B1+rms vs SAR?
    • Personnel exposure?
    • Cell phones?
  • Gradient Fields >
    • Gradient safety overview
    • Acoustic noise?
    • Nerve stimulation?
    • Gradient vs RF heating?
  • Safety: Neurological >
    • Aneurysm coils/clips?
    • Shunts/drains?
    • Pressure monitors/bolts?
    • Deep brain stimulators?
    • Spinal cord stimulators?
    • Vagal nerve stimulators?
    • Cranial electrodes?
    • Carotid clamps?
    • Peripheral stimulators?
    • Epidural catheters?
  • Safety: Head & Neck >
    • Additional orbit safety?
    • Cochlear Implants?
    • Bone conduction implants?
    • Other ear implants?
    • Dental/facial implants?
    • ET tubes & airways?
  • Safety: Chest & Vascular >
    • Breast tissue expanders?
    • Breast biopsy markers?
    • Airway stents/valves/coils?
    • Respiratory stimulators?
    • Ports/vascular access?
    • Swan-Ganz catheters?
    • IVC filters?
    • Implanted infusion pumps?
    • Insulin pumps & CGMs?
    • Vascular stents/grafts?
    • Sternal wires/implants?
  • Safety: Cardiac >
    • Pacemaker dangers?
    • Pacemaker terminology?
    • New/'Safe" Pacemakers?
    • Old/Legacy Pacemakers?
    • Violating the conditions?
    • Epicardial pacers/leads?
    • Cardiac monitors?
    • Heart valves?
    • Miscellaneous CV devices?
  • Safety: Abdominal >
    • PIllCam and capsules?
    • Gastric pacemakers?
    • Other GI devices?
    • Contraceptive devices?
    • Foley catheters?
    • Incontinence devices?
    • Penile Implants?
    • Sacral nerve stimulators?
    • GU stents and other?
  • Safety: Orthopedic >
    • Orthopedic hardware?
    • External fixators?
    • Traction and halos?
    • Bone stimulators?
    • Magnetic rods?
  • Spin >
    • What is spin?
    • Why I = ½, 1, etc?
    • Proton = nucleus = spin?
    • Predict nuclear spin (I)?
    • Magnetic dipole moment?
    • Gyromagnetic ratio (γ)?
    • "Spin" vs "Spin state"?
    • Energy splitting?
    • Fall to lowest state?
    • Quantum "reality"?
  • Precession >
    • Why precession?
    • Who was Larmor?
    • Energy for precession?
    • Chemical shift?
    • Net magnetization (M)?
    • Does M instantly appear?
    • Does M also precess?
    • Does precession = NMR?
  • Resonance >
    • MR vs MRI vs NMR?
    • Who discovered NMR?
    • How does B1 tip M?
    • Why at Larmor frequency?
    • What is flip angle?
    • Spins precess after 180°?
    • Phase coherence?
    • Release of RF energy?
    • Rotating frame?
    • Off-resonance?
    • Adiabatic excitation?
    • Adiabatic pulses?
  • Relaxation - Physics >
    • Bloch equations?
    • What is T1?
    • What is T2?
    • Relaxation rate vs time?
    • Why is T1 > T2?
    • T2 vs T2*?
    • Causes of Relaxation?
    • Dipole-dipole interactions?
    • Chemical Exchange?
    • Spin-Spin interactions?
    • Macromolecule effects?
    • Which H's produce signal?
    • "Invisible" protons?
    • Magnetization Transfer?
    • Bo effect on T1 & T2?
    • How to predict T1 & T2?
  • Relaxation - Clincial >
    • T1 bright? - fat
    • T1 bright? - other oils
    • T1 bright? - cholesterol
    • T1 bright? - calcifications
    • T1 bright? - meconium
    • T1 bright? - melanin
    • T1 bright? - protein/mucin
    • T1 bright? - myelin
    • Magic angle?
    • MT Imaging/Contrast?
  • MR Signals >
    • Origin of MR signal?
    • Free Induction Decay?
    • Gradient echo?
    • TR and TE?
    • Spin echo?
    • 90°-90° Hahn Echo?
    • Stimulated echoes?
    • STEs for imaging?
    • 4 or more RF-pulses?
    • Partial flip angles?
    • How is signal higher?
    • Optimal flip angle?
  • Spin Echo >
    • SE vs Multi-SE vs FSE?
    • Image contrast: TR/TE?
    • Opposite effects ↑T1 ↑T2?
    • Meaning of weighting?
    • Does SE correct for T2?
    • Effect of 180° on Mz?
    • Direction of 180° pulse?
  • Inversion Recovery >
    • What is IR?
    • Why use IR?
    • Phase-sensitive IR?
    • Why not PSIR always?
    • Choice of IR parameters?
    • TI to null a tissue?
    • STIR?
    • T1-FLAIR
    • T2-FLAIR?
    • IR-prepped sequences?
    • Double IR?
  • Gradient Echo >
    • GRE vs SE?
    • Multi-echo GRE?
    • Types of GRE sequences?
    • Commercial Acronyms?
    • Spoiling - what and how?
    • Spoiled-GRE parameters?
    • Spoiled for T1W only?
    • What is SSFP?
    • GRASS/FISP: how?
    • GRASS/FISP: parameters?
    • GRASS vs MPGR?
    • PSIF vs FISP?
    • True FISP/FIESTA?
    • FIESTA v FIESTA-C?
    • DESS?
    • MERGE/MEDIC?
    • GRASE?
    • MP-RAGE v MR2RAGE?
  • Susceptibility Imaging >
    • What is susceptibility (χ)?
    • What's wrong with GRE?
    • Making an SW image?
    • Phase of blood v Ca++?
    • Quantitative susceptibility?
  • Diffusion: Basic >
    • What is diffusion?
    • Iso-/Anisotropic diffusion?
    • "Apparent" diffusion?
    • Making a DW image?
    • What is the b-value?
    • b0 vs b50?
    • Trace vs ADC map?
    • Light/dark reversal?
    • T2 "shine through"?
    • Exponential ADC?
    • T2 "black-out"?
    • DWI bright causes?
  • Diffusion: Advanced >
    • Diffusion Tensor?
    • DTI (tensor imaging)?
    • Whole body DWI?
    • Readout-segmented DWI?
    • Small FOV DWI?
    • IVIM?
    • Diffusion Kurtosis?
  • Fat-Water Imaging >
    • Fat & Water properties?
    • F-W chemical shift?
    • In-phase/out-of-phase?
    • Best method?
    • Dixon method?
    • "Fat-sat" pulses?
    • Water excitation?
    • STIR?
    • SPIR?
    • SPAIR v SPIR?
    • SPIR/SPAIR v STIR?
  • From Signals to Images >
    • Phase v frequency?
    • Angular frequency (ω)?
    • Signal squiggles?
    • Real v Imaginary?
    • Fourier Transform (FT)?
    • What are 2D- & 3D-FTs?
    • Who invented MRI?
    • How to locate signals?
  • Frequency Encoding >
    • Frequency encoding?
    • Receiver bandwidth?
    • Narrow bandwidth?
    • Slice-selective excitation?
    • SS gradient lobes?
    • Cross-talk?
    • Frequency encode all?
    • Mixing of slices?
    • Two slices at once?
    • Simultaneous Multi-Slice?
  • Phase Encoding >
    • Phase-encoding gradient?
    • Single PE step?
    • What is phase-encoding?
    • PE and FE together?
    • 2DFT reconstruction?
    • Choosing PE/FE direction?
  • Performing an MR Scan >
    • What are the steps?
    • Automatic prescan?
    • Routine shimming?
    • Coil tuning/matching?
    • Center frequency?
    • Transmitter gain?
    • Receiver gain?
    • Dummy cycles?
    • Where's my data?
    • MR Tech qualifications?
  • Image Quality Control >
    • Who regulates MRI?
    • Who accredits?
    • Mandatory accreditation?
    • Routine quality control?
    • MR phantoms?
    • Geometric accuracy?
    • Image uniformity?
    • Slice parameters?
    • Image resolution?
    • Signal-to-noise?
    • Ghosting?
  • K-space (Basic) >
    • What is k-space?
    • Parts of k-space?
    • What does "k" stand for?
    • Spatial frequencies?
    • Locations in k-space?
    • Data for k-space?
    • Why signal ↔ k-space?
    • Spin-warp imaging?
    • Big spot in middle?
    • K-space trajectories?
    • Radial sampling?
  • K-space (Advanced) >
    • K-space grid?
    • Negative frequencies?
    • Field-of-view (FOV)
    • Rectangular FOV?
    • Partial Fourier?
    • Phase symmetry?
    • Read symmetry?
    • Why not use both?
    • ZIP?
  • Rapid Imaging (FSE &EPI) >
    • What is FSE/TSE?
    • FSE parameters?
    • Bright Fat?
    • Other FSE differences?
    • Dual-echo FSE?
    • Driven equilibrium?
    • Reduced flip angle FSE?
    • Hyperechoes?
    • SPACE/CUBE/VISTA?
    • Echo-planar imaging?
    • HASTE/SS-FSE?
  • Parallel Imaging (PI) >
    • What is PI?
    • How is PI different?
    • PI coils and sequences?
    • Why and when to use?
    • Two types of PI?
    • SENSE/ASSET?
    • GRAPPA/ARC?
    • CAIPIRINHA?
    • Compressed sensing?
    • Noise in PI?
    • Artifacts in PI?
  • Contrast Agents: Physics >
    • Why Gadolinium?
    • Paramagnetic relaxation?
    • What is relaxivity?
    • Why does Gd shorten T1?
    • Does Gd affect T2?
    • Gd & field strength?
    • Best T1-pulse sequence?
    • Triple dose and MT?
    • Dynamic CE imaging?
    • Gadolinium on CT?
  • Contrast Agents: Clinical >
    • So many Gd agents!
    • Important properties?
    • Ionic v non-ionic?
    • Intra-articular/thecal Gd?
    • Gd liver agents (Eovist)?
    • Mn agents (Teslascan)?
    • Feridex & Liver Agents?
    • Lymph node agents?
    • Ferumoxytol?
    • Blood pool (Ablavar)?
    • Bowel contrast agents?
  • Contrast Agents: Safety >
    • Gadolinium safety?
    • Allergic reactions?
    • Renal toxicity?
    • What is NSF?
    • NSF by agent?
    • Informed consent for Gd?
    • Gd protocol?
    • Is Gd safe in infants?
    • Reduced dose in infants?
    • Gd in breast milk?
    • Gd in pregnancy?
    • Gd accumulation?
    • Gd deposition disease?
  • Flow effects in MRI >
    • Defining flow?
    • Expected velocities?
    • Laminar v turbulent?
    • Predicting MR of flow?
    • Time-of-flight effects?
    • Spin phase effects?
    • Flow void?
    • Why GRE ↑ flow signal?
    • Slow flow v thrombus?
    • Even-echo rephasing?
    • Flow-compensation?
    • Flow misregistration?
  • MR Angiography - I >
    • MRA methods?
    • Dark vs bright blood?
    • Time-of-Flight (TOF) MRA?
    • 2D vs 3D MRA?
    • MRA parameters?
    • Magnetization Transfer?
    • Ramped flip angle?
    • MOTSA?
    • Fat-suppressed MRA?
    • TOF MRA Artifacts?
    • Phase-contrast MRA?
    • What is VENC?
    • Measuring flow?
    • 4D Flow Imaging?
    • How accurate?
  • MR Angiography - II >
    • Gated 3D FSE MRA?
    • 3D FSE MRA parameters?
    • SSFP MRA?
    • Inflow-enhanced SSFP?
    • MRA with ASL?
    • Other MRA methods?
    • Contrast-enhanced MRA?
    • Timing the bolus?
    • View ordering in MRA?
    • Bolus chasing?
    • TRICKS or TWIST?
    • CE-MRA artifacts?
  • Cardiac I - Intro/Anatomy >
    • Cardiac protocols?
    • Patient prep?
    • EKG problems?
    • Magnet changes EKG?
    • Gating v triggering?
    • Gating parameters?
    • Heart navigators?
    • Dark blood/Double IR?
    • Why not single IR?
    • Triple IR?
    • Polar plots?
    • Coronary artery MRA?
  • Cardiac II - Function >
    • Beating heart movies?
    • Cine parameters?
    • Real-time cine?
    • Ventricular function?
    • Tagging/SPAMM?
    • Perfusion: why and how?
    • 1st pass perfusion?
    • Quantifying perfusion?
    • Dark rim artifact
  • Cardiac III - Viability >
    • Gd enhancement?
    • TI to null myocardium?
    • PS (phase-sensitive) IR?
    • Wideband LGE?
    • T1 mapping?
    • Iron/T2*-mapping?
    • Edema/T2-mapping?
    • Why/how stress test?
    • Stess drugs/agents?
    • Stress consent form?
  • Tissue-related artifacts >
    • Chemical shift artifact?
    • Chemical shift in phase?
    • Reducing chemical shift?
    • Chemical Shift 2nd Kind?
    • In-phase/out-of phase?
    • IR bounce point?
    • Susceptibility artifact?
    • Metal suppression?
    • Dielectric effect?
    • Dielectric Pads?
  • Motion-related artifacts >
    • Why discrete ghosts?
    • Motion artifact direction?
    • Reducing motion artifacts?
    • Saturation pulses?
    • Gating methods?
    • Respiratory comp?
    • Navigator echoes?
    • PROPELLER/BLADE?
  • Technique-related artifacts >
    • Partial volume effects?
    • Slice overlap?
    • Aliasing?
    • Wrap-around artifact?
    • Eliminate wrap-around?
    • Phase oversampling?
    • Frequency wrap-around?
    • Spiral/radial artifacts?
    • Gibbs artifact?
    • Nyquist (N/2) ghosts?
    • Zipper artifact?
    • Data artifacts?
    • Surface coil flare?
    • MRA Artifacts (TOF)?
    • MRA artifacts (CE)?
  • Perfusion I: Intro & DSC >
    • Measuring perfusion?
    • Meaning of CBF, MTT etc?
    • DSC v DCE v ASL?
    • How to perform DSC?
    • Bolus Gd effect?
    • T1 effects on DSC?
    • DSC recirculation?
    • DSC curve analysis?
    • DSC signal v [Gd]
    • Arterial input (AIF)?
    • Quantitative DSC?
  • Perfusion II: DCE >
    • What is DCE?
    • How is DCE performed?
    • How is DCE analyzed?
    • Breast DCE?
    • DCE signal v [Gd]
    • DCE tissue parmeters?
    • Parameters to images?
    • K-trans = permeability?
    • Utility of DCE?
  • Perfusion III: ASL >
    • What is ASL?
    • ASL methods overview?
    • CASL?
    • PASL?
    • pCASL?
    • ASL parameters?
    • ASL artifacts?
    • Gadolinium and ASL?
    • Vascular color maps?
    • Quantifying flow?
  • Functional MRI/BOLD - I >
    • Who invented fMRI?
    • How does fMRI work?
    • BOLD contrast?
    • Why does BOLD ↑ signal?
    • Does BOLD=brain activity?
    • BOLD pulse sequences?
    • fMRI Paradigm design?
    • Why "on-off" comparison?
    • Motor paradigms?
    • Visual?
    • Language?
  • Functional MRI/BOLD - II >
    • Process/analyze fMRI?
    • Best fMRI software?
    • Data pre-processing?
    • Registration/normalization?
    • fMRI statistical analysis?
    • General Linear Model?
    • Activation "blobs"?
    • False activation?
    • Resting state fMRI?
    • Analyze RS-fMRI?
    • Network/Graphs?
    • fMRI at 7T?
    • Mind reading/Lie detector?
    • fMRI critique?
  • MRS I - Basics >
    • MRI vs MRS?
    • Spectra vs images?
    • Chemical shift (δ)?
    • Measuring δ?
    • Backward δ scale?
    • Predicting δ?
    • Size/shapes of peaks?
    • Splitting of peaks?
    • Localization methods?
    • Single v multi-voxel?
    • PRESS?
    • STEAM?
    • ISIS?
    • CSI?
  • MRS II - Clinical ¹H MRS >
    • How-to: brain MRS?
    • Water suppression?
    • Fat suppression?
    • Normal brain spectra?
    • Choice of TR/TE/etc?
    • Hunter's angle?
    • Lactate inversion?
    • Metabolite mapping?
    • Metabolite quantitation?
    • Breast MRS?
    • Gd effect on MRS?
    • How-to: prostate MRS?
    • Prostate spectra?
    • Muscle ¹H-MRS?
    • Liver ¹H-MRS?
    • MRS artifacts?
  • MRS III - Multi-nuclear >
    • Other nuclei?
    • Why phosphorus?
    • How-to: ³¹P MRS
    • Normal ³¹P spectra?
    • Organ differences?
    • ³¹P measurements?
    • Decoupling?
    • NOE?
    • Carbon MRS?
    • Sodium imaging?
    • Xenon imaging?
  • AI Part I: Basics >
    • Artificial Intelligence (AI)?
    • What is a neural network?
    • Machine Learning (ML)?
    • Shallow v Deep ML?
    • Shallow networks?
    • Deep network types?
    • Data prep and fitting?
    • Back-Propagation?
    • DL 'Playground'?
  • AI Part 2: Advanced >
    • What is convolution?
    • Convolutional Network?
    • Softmax?
    • Upsampling?
    • Limitations/Problems of AI?
    • Is the Singularity near?
  • AI Part 3: Image processing >
    • AI in clinical MRI?
    • Super-resolution?
  • MRI of Hemorrhage >
    • Hematoma overview?
    • Types of Hemoglobin?
    • Hyperacute/Oxy-Hb?
    • Acute/Deoxy-Hb?
    • Subacute/Met-Hb?
    • Deoxy-Hb v Met-Hb?
    • Extracellular met-Hb?
    • Chronic hematomas?
    • Hemichromes?
    • Ferritin/Hemosiderin?
    • Subarachnoid blood?
    • Blood at lower fields?
  • T2 cartilage mapping
  • MR Elastography?
  • Synthetic MRI?
  • Amide Proton Transfer?
  • MR thermography?
  • Electric Properties Imaging?
  • Copyright Issues
  • Legal Disclaimers
  • Magnets & Scanners Quiz
  • Safety & Screening Quiz
  • NMR Phenomenon Quiz
  • Pulse Sequences Quiz
  • Making an Image Quiz
  • K-space & Rapid Quiz
  • Contrast & Blood Quiz
  • Cardiovascular & MRA Quiz