Use of Quantitative DCE

When might quantitative DCE imaging be useful? Can you give some examples?  
Picture

Frankly, I am not enthusiastic about the widespread clinical applicability of quantitative DCE at its present level of development. Looking critically at the existing literature, the reader will find significant site-to-site variability of measured parameters. When disparate tumors (such as lymphomas vs glioblastomas) are compared, statistically significant differences in variables like Ktrans and ve have been noted, but the overlap between groups makes these distinctions worthless in practice. In most of the brain tumor cases we have studied to date, quantitative DCE simply reinforces diagnostic probabilities obtainable by visual inspection of time-intensity contrast uptake curves or parameters obtainable by conventional, DSC, or DW imaging. If you do wish to believe, and want to see why some think quantitative DCE imaging has a bright future, read on. . .

The pathophysiological basis of DCE imaging lies in the (imperfect) concept that more malignant/aggressive neoplasms have greater degrees of neovascularity. To supply nutrients to their rapidly proliferating cells, blood flow and blood volumes typically increase. The fraction of tissue occupied by blood vessels and plasma vp may also increase. Tumor blood vessels have increased surface areas and are also inherently leaky. Thus their permeabiltity-surface area products are elevated, and we would expect their Ktrans values to be high as well. Additionally, malignant tumors often have regions of necrosis producing in an increased extracellular space fraction (ve) compared to more benign tumors.

One of the principal concepts driving oncologists to push for quantitative DCE imaging is its potential to predict and monitor tumor response. A number of new chemotherapeutic agents such as bevacizumab and cediranib are directed against vascular endothelial growth factor (VEGF). These agents are not directly cytotoxic against cancer cells, and hence do not result in immediate shrinkage of tumors. However, by reducing vascular permeability they affect Ktrans and vp, thus presenting a potentially quantifiable method to assess effectiveness of therapy. For similar reasons, DCE parameters hold promise to distinguish progression from pseudoprogression and recurrent tumor vs radionecrosis. Two of our cases showing application of DCE imaging to brain neoplasms after therapy are presented below.

Picture
↑ enhancement
Picture
↑ Ktrans
Picture
↑ ve
Picture
↑ vp
Picture
10 mos later

Above. DCE study positive/predictive for recurrent glioma. Contrast enhancement with elevations of Ktrans, ve, and vp. Follow-up at 10 months shows marked progression of tumor.

Below. Equivocal DCE study in a treated glioma. Contrast enhancement with marked elevation of Ktrans and ve noted, as well as a slight elevation of vp. Is this viable tumor? Follow-up 3 years later shows treatment related changes only, no tumor progression or recurrence.

Picture
↑ enhancement
Picture
↑ Ktrans
Picture
↑ ve
Picture
slight ↑ vp
Picture
3 year followup
Considerable interest over the last ten years has been directed toward the use of DCE to detect very subtle areas of increased blood-brain-barrier permeability not visible to the eye. This may be important in explaining certain mechanisms in Alzheimer dementia, as well as demyelinating and infectious/inflammatory disorders of the nervous system.  Even here, quantitative accuracy of such methods is being questioned.

Advanced Discussion (show/hide)»

No supplementary material yet. Check back soon!

References
     Anzalone N, Castellano A, Cadioli M, et al. Brain gliomas: multicenter standardized assessment of dynamic contrast-enhanced and dynamic susceptibility contrast MR images. Radiology 2018; (in press)(recent paper showing good results using DCE and DSC to predict grade of brain tumors)
    Beuzit L, Eliat P-A, Brun V, et al. Dynamic contrast-enhanced MRI: study of inter-software accuracy and reproducibility using simulated and clinical data. J Magn Reson Imaging 2016; 43:1288-1300. (Poor agreement and reproducibility testing 5 commercial DCE-MRI software packages).
    Choi HS, KIm AH, Ahn SS, et al. Glioma grading capability: comparisons among parameters from dynamic contrast-enhanced MRI and ADC value on DWI. Korean J Radiol 2013; 14:487-92. (sounds great, but remember that many of the low-grade gliomas enhance weakly, if at all, so DCE parameters will reflect this.)
     Chikui T, Obara M, Simonetti AW, et al. The principal (sp) of dynamic contrast enhanced MRI, the method of pharmacokinetic analysis, and its application in the head and neck region. Int J Dentistry 2012; doi:10.1155/2012/480659
     Fatterpekar GM, Galheigo D, Naryana A, et al. Treatment-related change versus tumor recurrence in high-grade gliomas: a diagnostic conundrum — use of dynamic susceptibility contrast-enhanced (DSC) perfusion MRI. Am J Roentgenol 2012; 198:19-26. (acutally about DSC, not DCE, but gives an excellent description of the pathology of treatment-related changes in brain tumors).
     Kickingereder P, Sahm F, Wiestler B, et al. Evaluation of microvascular permeability with dynamic contrast-enhanced MRI for the differentiation of primary CNS lymphoma and glioblastoma: radiologic-pathologic correlation. AJNR Am J Neuroradiol 2014; 35:1503-1508. (some statistical differences in DCE parameters noted, but significant overlap between the two diseases in my opinion makes the technique worthless.)
     Heye AK, Culling RD, del C Valdés Hernández M, et al. Assessment of blood-brain barrier disruption using dynamic contrast-enhanced MRI. A systematic review. NeuroImage: Clinical 2014; 6:262-274.  [DOI LINK]
     Manning C, Stringer M, Dickie B, et al. Sources of systematic error in DCE-MRI estimation of low-level blood-brain barrier leakage. Magn Reson Med 2021; (in press)  [DOI LINK]  (Casts doubt on accuracy of Patlak model for measuring BBB disruption)
     Zhang N, Zhang L, Qui B, et al. Correlation of volume transfer coefficient Ktrans with histopathologic grades of gliomas. J Magn Reson imaging 2012; 36:355-363.

Related Questions
     How are quantitative DCE parameters extracted from the signal data?  

←  Previous Question
Next Question  →
↑ Complete List of Questions ↑
© 2024 AD Elster, ELSTER LLC
All rights reserved.   
MRIquestions.com - Home
Donate
Please help keep this site free for everyone in the world!
  • Basic Electromagnetism >
    • What causes magnetism?
    • What is a Tesla?
    • Who was Tesla?
    • What is a Gauss?
    • How strong is 3.0T?
    • What is a gradient?
    • Aren't gradients coils?
    • What is susceptibility?
    • How to levitate a frog?
    • What is ferromagnetism?
    • Superparamagnetism?
  • Magnets - Part I >
    • Types of magnets?
    • Brands of scanners?
    • Which way does field point?
    • Which is the north pole?
    • Low v mid v high field?
    • Advantages to low-field?
    • Disadvantages?
    • What is homogeneity?
    • Why homogeneity?
    • Why shimming?
    • Passive shimming?
    • Active shimming?
  • Magnets - Part II >
    • Superconductivity?
    • Perpetual motion?
    • How to ramp?
    • Superconductive design?
    • Room Temp supercon?
    • Liquid helium use?
    • What is a quench?
    • Is field ever turned off?
    • Emergency stop button?
  • Gradients >
    • Gradient coils?
    • How do z-gradients work?
    • X- and Y- gradients?
    • Open scanner gradients?
    • Eddy current problems?
    • Active shielded gradients?
    • Active shield confusion?
    • What is pre-emphasis?
    • Gradient heating?
    • Gradient specifications?
    • Gradient linearity?
  • RF & Coils >
    • Many kinds of coils?
    • Radiofrequency waves?
    • Phase v frequency?
    • RF Coil function(s)?
    • RF-transmit coils?
    • LP vs CP (Quadrature)?
    • Multi-transmit RF?
    • Receive-only coils?
    • Array coils?
    • AIR Coils?
  • Site Planning >
    • MR system layout?
    • What are fringe fields?
    • How to reduce fringe?
    • Magnetic shielding?
    • Need for vibration testing?
    • What's that noise?
    • Why RF Shielding?
    • Wires/tubes thru wall?
  • Overview >
    • ACR Safety Zones?
    • MR safety screening?
    • Incomplete screening?
    • Passive v active implants?
    • Conditional implants?
    • Common safety issues?
    • Projectiles?
    • Metal detectors?
    • Pregnant patients?
    • Postop, ER & ICU patients?
    • Temperature monitoring?
    • Orbital foreign bodies?
    • Bullets and shrapnel?
  • Static Fields >
    • "Dangerous" metals?
    • "Safe" metals?
    • Magnetizing metal?
    • Object shape?
    • Forces on metal?
    • Most dangerous place?
    • Force/torque testing?
    • Static field bioeffects?
    • Dizziness/Vertigo?
    • Flickering lights?
    • Metallic taste?
  • RF Fields >
    • RF safety overview?
    • RF biological effects?
    • What is SAR?
    • SAR limits?
    • Operating modes?
    • How to reduce SAR?
    • RF burns?
    • Estimate implant heating?
    • SED vs SAR?
    • B1+rms vs SAR?
    • Personnel exposure?
    • Cell phones?
  • Gradient Fields >
    • Gradient safety overview
    • Acoustic noise?
    • Nerve stimulation?
    • Gradient vs RF heating?
  • Safety: Neurological >
    • Aneurysm coils/clips?
    • Shunts/drains?
    • Pressure monitors/bolts?
    • Deep brain stimulators?
    • Spinal cord stimulators?
    • Vagal nerve stimulators?
    • Cranial electrodes?
    • Carotid clamps?
    • Peripheral stimulators?
    • Epidural catheters?
  • Safety: Head & Neck >
    • Additional orbit safety?
    • Cochlear Implants?
    • Bone conduction implants?
    • Other ear implants?
    • Dental/facial implants?
    • ET tubes & airways?
  • Safety: Chest & Vascular >
    • Breast tissue expanders?
    • Breast biopsy markers?
    • Airway stents/valves/coils?
    • Respiratory stimulators?
    • Ports/vascular access?
    • Swan-Ganz catheters?
    • IVC filters?
    • Implanted infusion pumps?
    • Insulin pumps & CGMs?
    • Vascular stents/grafts?
    • Sternal wires/implants?
  • Safety: Cardiac >
    • Pacemaker dangers?
    • Pacemaker terminology?
    • New/'Safe" Pacemakers?
    • Old/Legacy Pacemakers?
    • Violating the conditions?
    • Epicardial pacers/leads?
    • Cardiac monitors?
    • Heart valves?
    • Miscellaneous CV devices?
  • Safety: Abdominal >
    • PIllCam and capsules?
    • Gastric pacemakers?
    • Other GI devices?
    • Contraceptive devices?
    • Foley catheters?
    • Incontinence devices?
    • Penile Implants?
    • Sacral nerve stimulators?
    • GU stents and other?
  • Safety: Orthopedic >
    • Orthopedic hardware?
    • External fixators?
    • Traction and halos?
    • Bone stimulators?
    • Magnetic rods?
  • Spin >
    • What is spin?
    • Why I = ½, 1, etc?
    • Proton = nucleus = spin?
    • Predict nuclear spin (I)?
    • Magnetic dipole moment?
    • Gyromagnetic ratio (γ)?
    • "Spin" vs "Spin state"?
    • Energy splitting?
    • Fall to lowest state?
    • Quantum "reality"?
  • Precession >
    • Why precession?
    • Who was Larmor?
    • Energy for precession?
    • Chemical shift?
    • Net magnetization (M)?
    • Does M instantly appear?
    • Does M also precess?
    • Does precession = NMR?
  • Resonance >
    • MR vs MRI vs NMR?
    • Who discovered NMR?
    • How does B1 tip M?
    • Why at Larmor frequency?
    • What is flip angle?
    • Spins precess after 180°?
    • Phase coherence?
    • Release of RF energy?
    • Rotating frame?
    • Off-resonance?
    • Adiabatic excitation?
    • Adiabatic pulses?
  • Relaxation - Physics >
    • Bloch equations?
    • What is T1?
    • What is T2?
    • Relaxation rate vs time?
    • Why is T1 > T2?
    • T2 vs T2*?
    • Causes of Relaxation?
    • Dipole-dipole interactions?
    • Chemical Exchange?
    • Spin-Spin interactions?
    • Macromolecule effects?
    • Which H's produce signal?
    • "Invisible" protons?
    • Magnetization Transfer?
    • Bo effect on T1 & T2?
    • How to predict T1 & T2?
  • Relaxation - Clincial >
    • T1 bright? - fat
    • T1 bright? - other oils
    • T1 bright? - cholesterol
    • T1 bright? - calcifications
    • T1 bright? - meconium
    • T1 bright? - melanin
    • T1 bright? - protein/mucin
    • T1 bright? - myelin
    • Magic angle?
    • MT Imaging/Contrast?
  • MR Signals >
    • Origin of MR signal?
    • Free Induction Decay?
    • Gradient echo?
    • TR and TE?
    • Spin echo?
    • 90°-90° Hahn Echo?
    • Stimulated echoes?
    • STEs for imaging?
    • 4 or more RF-pulses?
    • Partial flip angles?
    • How is signal higher?
    • Optimal flip angle?
  • Spin Echo >
    • SE vs Multi-SE vs FSE?
    • Image contrast: TR/TE?
    • Opposite effects ↑T1 ↑T2?
    • Meaning of weighting?
    • Does SE correct for T2?
    • Effect of 180° on Mz?
    • Direction of 180° pulse?
  • Inversion Recovery >
    • What is IR?
    • Why use IR?
    • Phase-sensitive IR?
    • Why not PSIR always?
    • Choice of IR parameters?
    • TI to null a tissue?
    • STIR?
    • T1-FLAIR
    • T2-FLAIR?
    • IR-prepped sequences?
    • Double IR?
  • Gradient Echo >
    • GRE vs SE?
    • Multi-echo GRE?
    • Types of GRE sequences?
    • Commercial Acronyms?
    • Spoiling - what and how?
    • Spoiled-GRE parameters?
    • Spoiled for T1W only?
    • What is SSFP?
    • GRASS/FISP: how?
    • GRASS/FISP: parameters?
    • GRASS vs MPGR?
    • PSIF vs FISP?
    • True FISP/FIESTA?
    • FIESTA v FIESTA-C?
    • DESS?
    • MERGE/MEDIC?
    • GRASE?
    • MP-RAGE v MR2RAGE?
  • Susceptibility Imaging >
    • What is susceptibility (χ)?
    • What's wrong with GRE?
    • Making an SW image?
    • Phase of blood v Ca++?
    • Quantitative susceptibility?
  • Diffusion: Basic >
    • What is diffusion?
    • Iso-/Anisotropic diffusion?
    • "Apparent" diffusion?
    • Making a DW image?
    • What is the b-value?
    • b0 vs b50?
    • Trace vs ADC map?
    • Light/dark reversal?
    • T2 "shine through"?
    • Exponential ADC?
    • T2 "black-out"?
    • DWI bright causes?
  • Diffusion: Advanced >
    • Diffusion Tensor?
    • DTI (tensor imaging)?
    • Whole body DWI?
    • Readout-segmented DWI?
    • Small FOV DWI?
    • IVIM?
    • Diffusion Kurtosis?
  • Fat-Water Imaging >
    • Fat & Water properties?
    • F-W chemical shift?
    • In-phase/out-of-phase?
    • Best method?
    • Dixon method?
    • "Fat-sat" pulses?
    • Water excitation?
    • STIR?
    • SPIR?
    • SPAIR v SPIR?
    • SPIR/SPAIR v STIR?
  • From Signals to Images >
    • Phase v frequency?
    • Angular frequency (ω)?
    • Signal squiggles?
    • Real v Imaginary?
    • Fourier Transform (FT)?
    • What are 2D- & 3D-FTs?
    • Who invented MRI?
    • How to locate signals?
  • Frequency Encoding >
    • Frequency encoding?
    • Receiver bandwidth?
    • Narrow bandwidth?
    • Slice-selective excitation?
    • SS gradient lobes?
    • Cross-talk?
    • Frequency encode all?
    • Mixing of slices?
    • Two slices at once?
    • Simultaneous Multi-Slice?
  • Phase Encoding >
    • Phase-encoding gradient?
    • Single PE step?
    • What is phase-encoding?
    • PE and FE together?
    • 2DFT reconstruction?
    • Choosing PE/FE direction?
  • Performing an MR Scan >
    • What are the steps?
    • Automatic prescan?
    • Routine shimming?
    • Coil tuning/matching?
    • Center frequency?
    • Transmitter gain?
    • Receiver gain?
    • Dummy cycles?
    • Where's my data?
    • MR Tech qualifications?
  • Image Quality Control >
    • Who regulates MRI?
    • Who accredits?
    • Mandatory accreditation?
    • Routine quality control?
    • MR phantoms?
    • Geometric accuracy?
    • Image uniformity?
    • Slice parameters?
    • Image resolution?
    • Signal-to-noise?
    • Ghosting?
  • K-space (Basic) >
    • What is k-space?
    • Parts of k-space?
    • What does "k" stand for?
    • Spatial frequencies?
    • Locations in k-space?
    • Data for k-space?
    • Why signal ↔ k-space?
    • Spin-warp imaging?
    • Big spot in middle?
    • K-space trajectories?
    • Radial sampling?
  • K-space (Advanced) >
    • K-space grid?
    • Negative frequencies?
    • Field-of-view (FOV)
    • Rectangular FOV?
    • Partial Fourier?
    • Phase symmetry?
    • Read symmetry?
    • Why not use both?
    • ZIP?
  • Rapid Imaging (FSE &EPI) >
    • What is FSE/TSE?
    • FSE parameters?
    • Bright Fat?
    • Other FSE differences?
    • Dual-echo FSE?
    • Driven equilibrium?
    • Reduced flip angle FSE?
    • Hyperechoes?
    • SPACE/CUBE/VISTA?
    • Echo-planar imaging?
    • HASTE/SS-FSE?
  • Parallel Imaging (PI) >
    • What is PI?
    • How is PI different?
    • PI coils and sequences?
    • Why and when to use?
    • Two types of PI?
    • SENSE/ASSET?
    • GRAPPA/ARC?
    • CAIPIRINHA?
    • Compressed sensing?
    • Noise in PI?
    • Artifacts in PI?
  • Contrast Agents: Physics >
    • Why Gadolinium?
    • Paramagnetic relaxation?
    • What is relaxivity?
    • Why does Gd shorten T1?
    • Does Gd affect T2?
    • Gd & field strength?
    • Best T1-pulse sequence?
    • Triple dose and MT?
    • Dynamic CE imaging?
    • Gadolinium on CT?
  • Contrast Agents: Clinical >
    • So many Gd agents!
    • Important properties?
    • Ionic v non-ionic?
    • Intra-articular/thecal Gd?
    • Gd liver agents (Eovist)?
    • Mn agents (Teslascan)?
    • Feridex & Liver Agents?
    • Lymph node agents?
    • Ferumoxytol?
    • Blood pool (Ablavar)?
    • Bowel contrast agents?
  • Contrast Agents: Safety >
    • Gadolinium safety?
    • Allergic reactions?
    • Renal toxicity?
    • What is NSF?
    • NSF by agent?
    • Informed consent for Gd?
    • Gd protocol?
    • Is Gd safe in infants?
    • Reduced dose in infants?
    • Gd in breast milk?
    • Gd in pregnancy?
    • Gd accumulation?
    • Gd deposition disease?
  • Flow effects in MRI >
    • Defining flow?
    • Expected velocities?
    • Laminar v turbulent?
    • Predicting MR of flow?
    • Time-of-flight effects?
    • Spin phase effects?
    • Flow void?
    • Why GRE ↑ flow signal?
    • Slow flow v thrombus?
    • Even-echo rephasing?
    • Flow-compensation?
    • Flow misregistration?
  • MR Angiography - I >
    • MRA methods?
    • Dark vs bright blood?
    • Time-of-Flight (TOF) MRA?
    • 2D vs 3D MRA?
    • MRA parameters?
    • Magnetization Transfer?
    • Ramped flip angle?
    • MOTSA?
    • Fat-suppressed MRA?
    • TOF MRA Artifacts?
    • Phase-contrast MRA?
    • What is VENC?
    • Measuring flow?
    • 4D Flow Imaging?
    • How accurate?
  • MR Angiography - II >
    • Gated 3D FSE MRA?
    • 3D FSE MRA parameters?
    • SSFP MRA?
    • Inflow-enhanced SSFP?
    • MRA with ASL?
    • Other MRA methods?
    • Contrast-enhanced MRA?
    • Timing the bolus?
    • View ordering in MRA?
    • Bolus chasing?
    • TRICKS or TWIST?
    • CE-MRA artifacts?
  • Cardiac I - Intro/Anatomy >
    • Cardiac protocols?
    • Patient prep?
    • EKG problems?
    • Magnet changes EKG?
    • Gating v triggering?
    • Gating parameters?
    • Heart navigators?
    • Dark blood/Double IR?
    • Why not single IR?
    • Triple IR?
    • Polar plots?
    • Coronary artery MRA?
  • Cardiac II - Function >
    • Beating heart movies?
    • Cine parameters?
    • Real-time cine?
    • Ventricular function?
    • Tagging/SPAMM?
    • Perfusion: why and how?
    • 1st pass perfusion?
    • Quantifying perfusion?
    • Dark rim artifact
  • Cardiac III - Viability >
    • Gd enhancement?
    • TI to null myocardium?
    • PS (phase-sensitive) IR?
    • Wideband LGE?
    • T1 mapping?
    • Iron/T2*-mapping?
    • Edema/T2-mapping?
    • Why/how stress test?
    • Stess drugs/agents?
    • Stress consent form?
  • Tissue-related artifacts >
    • Chemical shift artifact?
    • Chemical shift in phase?
    • Reducing chemical shift?
    • Chemical Shift 2nd Kind?
    • In-phase/out-of phase?
    • IR bounce point?
    • Susceptibility artifact?
    • Metal suppression?
    • Dielectric effect?
    • Dielectric Pads?
  • Motion-related artifacts >
    • Why discrete ghosts?
    • Motion artifact direction?
    • Reducing motion artifacts?
    • Saturation pulses?
    • Gating methods?
    • Respiratory comp?
    • Navigator echoes?
    • PROPELLER/BLADE?
  • Technique-related artifacts >
    • Partial volume effects?
    • Slice overlap?
    • Aliasing?
    • Wrap-around artifact?
    • Eliminate wrap-around?
    • Phase oversampling?
    • Frequency wrap-around?
    • Spiral/radial artifacts?
    • Gibbs artifact?
    • Nyquist (N/2) ghosts?
    • Zipper artifact?
    • Data artifacts?
    • Surface coil flare?
    • MRA Artifacts (TOF)?
    • MRA artifacts (CE)?
  • Perfusion I: Intro & DSC >
    • Measuring perfusion?
    • Meaning of CBF, MTT etc?
    • DSC v DCE v ASL?
    • How to perform DSC?
    • Bolus Gd effect?
    • T1 effects on DSC?
    • DSC recirculation?
    • DSC curve analysis?
    • DSC signal v [Gd]
    • Arterial input (AIF)?
    • Quantitative DSC?
  • Perfusion II: DCE >
    • What is DCE?
    • How is DCE performed?
    • How is DCE analyzed?
    • Breast DCE?
    • DCE signal v [Gd]
    • DCE tissue parmeters?
    • Parameters to images?
    • K-trans = permeability?
    • Utility of DCE?
  • Perfusion III: ASL >
    • What is ASL?
    • ASL methods overview?
    • CASL?
    • PASL?
    • pCASL?
    • ASL parameters?
    • ASL artifacts?
    • Gadolinium and ASL?
    • Vascular color maps?
    • Quantifying flow?
  • Functional MRI/BOLD - I >
    • Who invented fMRI?
    • How does fMRI work?
    • BOLD contrast?
    • Why does BOLD ↑ signal?
    • Does BOLD=brain activity?
    • BOLD pulse sequences?
    • fMRI Paradigm design?
    • Why "on-off" comparison?
    • Motor paradigms?
    • Visual?
    • Language?
  • Functional MRI/BOLD - II >
    • Process/analyze fMRI?
    • Best fMRI software?
    • Data pre-processing?
    • Registration/normalization?
    • fMRI statistical analysis?
    • General Linear Model?
    • Activation "blobs"?
    • False activation?
    • Resting state fMRI?
    • Analyze RS-fMRI?
    • Network/Graphs?
    • fMRI at 7T?
    • Mind reading/Lie detector?
    • fMRI critique?
  • MRS I - Basics >
    • MRI vs MRS?
    • Spectra vs images?
    • Chemical shift (δ)?
    • Measuring δ?
    • Backward δ scale?
    • Predicting δ?
    • Size/shapes of peaks?
    • Splitting of peaks?
    • Localization methods?
    • Single v multi-voxel?
    • PRESS?
    • STEAM?
    • ISIS?
    • CSI?
  • MRS II - Clinical ¹H MRS >
    • How-to: brain MRS?
    • Water suppression?
    • Fat suppression?
    • Normal brain spectra?
    • Choice of TR/TE/etc?
    • Hunter's angle?
    • Lactate inversion?
    • Metabolite mapping?
    • Metabolite quantitation?
    • Breast MRS?
    • Gd effect on MRS?
    • How-to: prostate MRS?
    • Prostate spectra?
    • Muscle ¹H-MRS?
    • Liver ¹H-MRS?
    • MRS artifacts?
  • MRS III - Multi-nuclear >
    • Other nuclei?
    • Why phosphorus?
    • How-to: ³¹P MRS
    • Normal ³¹P spectra?
    • Organ differences?
    • ³¹P measurements?
    • Decoupling?
    • NOE?
    • Carbon MRS?
    • Sodium imaging?
    • Xenon imaging?
  • AI Part I: Basics >
    • Artificial Intelligence (AI)?
    • What is a neural network?
    • Machine Learning (ML)?
    • Shallow v Deep ML?
    • Shallow networks?
    • Deep network types?
    • Data prep and fitting?
    • Back-Propagation?
    • DL 'Playground'?
  • AI Part 2: Advanced >
    • What is convolution?
    • Convolutional Network?
    • Softmax?
    • Upsampling?
    • Limitations/Problems of AI?
    • Is the Singularity near?
  • AI Part 3: Image processing >
    • AI in clinical MRI?
    • Super-resolution?
  • MRI of Hemorrhage >
    • Hematoma overview?
    • Types of Hemoglobin?
    • Hyperacute/Oxy-Hb?
    • Acute/Deoxy-Hb?
    • Subacute/Met-Hb?
    • Deoxy-Hb v Met-Hb?
    • Extracellular met-Hb?
    • Chronic hematomas?
    • Hemichromes?
    • Ferritin/Hemosiderin?
    • Subarachnoid blood?
    • Blood at lower fields?
  • T2 cartilage mapping
  • MR Elastography?
  • Synthetic MRI?
  • Amide Proton Transfer?
  • MR thermography?
  • Electric Properties Imaging?
  • Copyright Issues
  • Legal Disclaimers
  • Magnets & Scanners Quiz
  • Safety & Screening Quiz
  • NMR Phenomenon Quiz
  • Pulse Sequences Quiz
  • Making an Image Quiz
  • K-space & Rapid Quiz
  • Contrast & Blood Quiz
  • Cardiovascular & MRA Quiz